Oligo Concentration Calculator from A₂₆₀ or Mass
Convert a NanoDrop reading or a weighed mass into molar concentration for any DNA oligonucleotide. The calculator derives sequence-specific extinction coefficient ε₂₆₀ and molecular weight, then applies the Beer–Lambert law.
Calculate oligo concentration
Paste your sequence and enter either an A₂₆₀ absorbance or a mass plus volume. The calculator reports µM, ng/µL, and conversion factors specific to the sequence.
Oligonucleotide sequence
A₂₆₀ reading
NanoDrop reading at 1 cm equivalent path length. Typical primer stocks: 0.4–3.0.
Path length
Standard cuvette: 1 cm. NanoDrop reports normalised to 1 cm by default.
Calculation
Useful conversions for this oligo
Beer–Lambert law for oligonucleotides
August Beer and Johann Heinrich Lambert worked out the linear absorbance relationship in the 19th century. Their law states that absorbance equals the molar extinction coefficient ε multiplied by concentration c and path length l: A = ε · c · l. Rearranging gives c = A / (ε · l), the working equation for concentration from any absorbance reading.
Nucleic acids absorb strongly at 260 nm because of π-π* electronic transitions in the aromatic purine and pyrimidine bases. Each base contributes a characteristic ε: adenine 15,200, thymine 8,400, guanine 12,010, cytosine 7,050 M⁻¹cm⁻¹. A naive sum overestimates the actual oligonucleotide ε because adjacent bases stack and shield each other from incident light, an effect called hypochromicity. The nearest-neighbor correction subtracts this stacking contribution by averaging dinucleotide-step extinctions.
A practical consequence: two oligos of the same length but different sequence can differ in ε by 20% or more. An AT-heavy 20-mer might have ε near 220,000, while a GC-heavy 20-mer of the same length stays closer to 170,000. The generic "1 OD = 33 µg/mL" rule, which assumes dsDNA, fails badly for short ssDNA. Sequence-specific ε is necessary for any quantitative work.
How NanoDrop and spectrophotometers measure A₂₆₀
Conventional UV-Vis spectrophotometers use 1 cm path-length quartz cuvettes with 50–200 µL sample volumes. The instrument compares transmitted light through the sample against a buffer blank and reports absorbance directly. NanoDrop instruments shrink the sample to 1–2 µL by exploiting surface tension between two optical pedestals, with a 0.05–1 mm variable path length. The displayed A₂₆₀ is normalised to 1 cm equivalent, so the same Beer–Lambert arithmetic applies regardless of which platform produced the reading.
Working concentrations sit between A₂₆₀ = 0.1 and 2.0 for accurate quantitation. Below 0.1, instrument noise dominates and readings carry 10–20% relative error. Above 2.0, the linear dynamic range breaks down because too little light reaches the detector. Dilute concentrated stocks 1:10 before reading if A₂₆₀ exceeds 2.0; the working dilution returns to the linear range and the result scales back by the dilution factor.
Worked examples
Example 1: NanoDrop reading of a primer stock
Sequence: 5′-ACCACAGTCCATGCCATCAC-3′ (20 nt). NanoDrop reads A₂₆₀ = 1.2 at 1 cm equivalent path.
Calculated ε₂₆₀ for this sequence ≈ 196,800 M⁻¹cm⁻¹ (from nearest-neighbor sum).
c = 1.2 / (196,800 × 1) = 6.1 × 10⁻⁶ M = 6.1 µM
Mass concentration: 6.1 µM × 6,135 g/mol ÷ 1000 = 37 ng/µL. This stock matches a typical resuspended PCR primer; dilute 1:60 in TE to reach a 100 nM working stock.
Example 2: Mass-based quantification
A 30-mer ssDNA antisense oligo. QC sheet reports 50 µg mass in 100 µL of TE buffer. Sequence-derived MW = 9,250 g/mol.
c = mass / (MW × V) = 50 × 10⁻⁶ g / (9,250 g/mol × 100 × 10⁻⁶ L) = 54 × 10⁻⁶ M = 54 µM
Total moles = 50 µg / 9,250 g/mol = 5.4 nmol. Verify by A₂₆₀ measurement and compare against this calculation. A discrepancy above 10% indicates incomplete resuspension or a yield reporting error.
When to use A₂₆₀ versus mass calculation
A₂₆₀ wins for routine bench quantification. The measurement takes 30 seconds, requires 1–2 µL of sample, and integrates over the full ε of the oligonucleotide. Resuspended stocks and working dilutions are quickest to verify this way. The downside: any UV-absorbing contaminant (modified bases, fluorophores, residual ammonia from synthesis cleavage) inflates the reading.
Mass-based calculation matters when you start from a known weighed amount, typically the vendor-reported yield on the QC sheet. This approach is independent of any spectroscopic interference but assumes the oligo is fully soluble and that the reported mass is accurate to within 5%. For modified oligos with fluorophores or other chromophores at 260 nm, mass-based is often more reliable than A₂₆₀.
For absolute accuracy, run both. A 10% agreement between the two methods confirms that the stock matches expectations. Larger discrepancies point to insoluble pellet residue, vendor yield miscalculation, or chromophore contamination.
Quality indicators in absorbance ratios
Modern spectrophotometers report A₂₆₀, A₂₈₀, and A₂₃₀ simultaneously. The ratios flag contamination types. A₂₆₀/A₂₈₀ near 1.8–2.0 indicates clean nucleic acid; below 1.7 suggests protein carryover (proteins absorb at 280 nm via tryptophan and tyrosine). A₂₆₀/A₂₃₀ above 2.0 confirms absence of guanidinium, phenol, or carbohydrate contamination from silica-column extraction.
Synthetic oligonucleotides shipped lyophilised typically score A₂₆₀/A₂₈₀ around 1.95–2.05 and A₂₆₀/A₂₃₀ above 2.2. Values outside these ranges point to a problem with the synthesis or with how you resuspended the pellet. For genomic DNA preparations the ratios diagnose nuclease purification quality; for oligonucleotides they diagnose synthesis cleavage and deprotection completeness.
Limitations and caveats
Sequence-derived ε₂₆₀ values assume only Watson–Crick A, T, G, and C residues at neutral pH. Modified bases such as inosine, 5-methylcytosine, locked nucleic acid, or 2′-O-methyl RNA shift the extinction coefficient by 2–10%. Fluorophore labels add discrete chromophore contributions that must be subtracted from the total absorbance before applying Beer–Lambert.
At low concentrations (below 10 nM), polypropylene tube adsorption removes a measurable fraction of oligo from solution. The A₂₆₀ reading reflects what stays in solution, not what was originally pipetted. Carrier nucleic acid (10 ng/µL tRNA) or low-binding tubes prevent this loss for sub-nanomolar work.
This calculator is intended for teaching and routine lab use. Pharmaceutical and clinical quantitation should use validated HPLC or LC-MS methods alongside spectroscopy. The IDT OligoAnalyzer and similar commercial tools cross-check ε values against independently calibrated reference standards.
Frequently asked questions
How do I convert A260 to oligo concentration?
Why is the 1 OD = 33 µg/mL rule inaccurate for short oligos?
What is the extinction coefficient and how is it calculated?
Why does my A260 reading differ from the QC sheet value?
How do I quantify a modified oligo (FAM, biotin, LNA)?
What buffer is best for A260 measurement?
Why is A260/A280 ratio reported alongside concentration?
References. Nearest-neighbor extinction coefficients follow Cantor, Warshaw, and Shapiro 1970. Spectrophotometric method overview available from NCBI.
Related Tools
Oligo Resuspension Calculator
Determine the diluent volume to dissolve a lyophilised oligo at your target concentration.
Open CalculatorOligo Dilution Calculator
Prepare a working dilution from a stock using C1V1 = C2V2.
Open CalculatorOligo Analyzer
Check Tm, hairpin, self-dimer, and thermodynamics for any DNA primer.
Open CalculatorGC Content Calculator
Measure GC content and base composition for DNA or RNA sequences.
Open Calculator